![]() |
| Module 3: Equipment for proton therapy delivery |
| Eric Shinohara MD, MSCI |
| The Abramson Cancer Center of the University of Pennsylvania |
|
|
Accelerators: Cyclotrons and SynchrotronsThe first step in generating a proton beam is to obtain a source of protons which can then be accelerated to energies sufficient for treatment. This can be done using hydrogen as the starting product and separating the hydrogen's electron from its proton by using an electrical field. Once protons have been generated, they must be accelerated such that the proton energy is sufficient to reach the distal edge of a tumor. Particle accelerators use an electrical field to accelerate protons and a magnet field to steer the charged particles. Linear accelerators are commonly used in photon therapy. In photon therapy, electrons are accelerated with a single pass through a series of electrical fields, and at the end of the beam line, the electrons are steered toward a target to generate photons, or alternatively to a scatter foil for electron therapy. Proton therapy requires cyclical particle accelerators which cause the particle to pass through the electrical field repeated times until they reach an energy sufficient for clinical use.
Beam Transport, Degraders, Range Modulation, and Current Modulation
The proton beam can be turned off at specific points during the revolution of the modulator wheel, which changes the width of the treatment field. This does have the disadvantage of increasing treatment times (as the beam is off for a portion of treatment) as well as making the system more complex. The beam current can also be varied precisely with the rotation of the modulator wheel, which can further reduce the number of modulator wheels needed; however, the complexity of the system is increased even further. Using this technique also requires a very sensitive current detector and changes in current which are very stable and linear.
Gantries and Inclined Beam Systems
The incline beam system uses two beams, a horizontal beam and a second beam which is angled to 30 degrees off of the vertical. These beams use a common isocenter and can be used together, in conjunction with a robotic patient positioner, to achieve a wide array of angles to treat the patient (Figure 8. Procure). There are also fixed beams which can only deliver protons in a single direction. These beams rely on the movement of the treatment table or chair around the beam to allow multiple angles to be treated.
NozzlesNozzles are used to deliver the protons to the patient and are comprised of multiple components (Figure 9. IBA). There are two main types of proton delivery systems used at present, passive scatter and scanning beams. Passive scattering will be discussed first.
Passive Scatter, Apertures, and CompensatorsIn a passive scatter system, the nozzle contains the above mentioned components including the scatter foils, ridge filter or modulator wheel, the aperture and the range compensator. In addition to spreading the proton field out to cover the depth of the tumor, the lateral aspects of the field also need to be expanded from the original thin pencil beam. A double scatter foil system can be used to broaden the proton beam laterally (green discs in figure 6). The first scatter foil used in this system is uniform and creates a Gaussian, or bell shaped, distribution of protons. However, this must then be made flat prior to reaching the patient. Hence, a second, non-uniform scatter foil is needed. The central protons are scattered to a greater degree compared with the protons in the lateral aspect of the beam, which acts to flatten the proton beam. Much like a flattening filter in photon therapy, any misalignment of the non-uniform, second scatter foil can create a skewed beam. In a single scatter foil system the single scatter foil causes a Gaussian distribution of proton energies, however, only the center portion, where the proton energies are within about 5% of the highest energy protons produced, can be used. This greatly limits the beam size. Additionally, the double scatter foil system allows less energy loss than the single scatter foil system. An aperture can then be used to further shape the lateral borders of the beam ('collimator' in figure 6). These are generally made of brass and can be quite heavy and difficult to manipulate. Both apertures and compensators are inserted into the 'snout' of the nozzle. There is currently a MLC system in development for protons at the University of Pennsylvania which would prevent the need for custom apertures. The MLC development process is described in greater detail here: Designing a Multileaf Collimator for Proton Therapy.
There are several disadvantages to the passive scatter system. Due to the need for an aperture and compensator, the proton beam must travel through several layers, creating more neutrons. Neutrons contamination delivers unwanted dose to the patient. As described above, the use of a compensator can shift the full dose into areas proximal to the tumor, which receive unwanted high doses. Passive scatter treatments also require many custom components to be made for each patient. Custom compensators and apertures must be made for each field used to treat a patient. The apertures become radioactive after treatment and need to be stored until they can be safely disposed. All of these components also need to have quality and assurance checks performed which can be time consuming. Spot Scanning SystemsThe second type of proton therapy which is just starting to be used in the United States is scanning beam proton therapy. Scanning beams use magnets to move the proton beam precisely, such that it can 'paint' the area that is to be treated. In a spot scanning system the nozzle contains the magnets needed to steer the proton beam. Once an area at a given depth has been treated, the energy of the proton can be changed and the next 'layer' can be painted. By repeating this, it is possible to treat the entire tumor. This technique allows greater conformality with shaping of the distal and proximal ends of the proton field. Scanning proton beams also allow the use of Intensity Modulated Proton Therapy (IMPT). With IMPT, multiple beams are used and a computer algorithm calculates the optimal arrangement of individual Bragg peaks needed to cover the volume. The sum of these Bragg peaks can provide a precise distribution of dose throughout the three- dimensional volume. Fewer neutrons are produced with scanning beams, as a compensator, scatter foil and aperture are not needed. The major disadvantage to the scanning beam is the greater complexity and longer treatment times due to the multiple 'layers' which must be 'painted'. There are also significant challenges to using scanning beams in areas of organ motion, as this technology is more susceptible to problems with motion. Dose DeterminationSimilar to photon-based therapy, the dose of radiation delivered to a patient with protons is measured in monitor units. These monitor units correspond to the known amount of charge collected in an ionization chamber present within the beam. The ion chamber plays the critical role of being the absolute reference monitor for the delivered dose. This also means that the dose prescription in gray must be converted to monitor units. The conversion factor needed to convert dose to monitor units is known as an output factor, and several models for output factor have been developed for photons. For protons, an actual measurement of the output for individual fields is required to accurately determine output factors. Numerous models, predominantly algorithms which use Monte Carlo calculations, to determine the output factor for a given treatment field are under development. Determination of the dose delivered by a proton beam is generally performed with an ionization chamber which has been specially designed for use with protons. Usually, ither a cylindrical ionization chamber or a plane-parallel chamber (for large depth dose gradients or narrow spread out Bragg peaks) is used. Scanning beams present additional challenges. The spot scanning beam requires close monitoring to ensure that each spot is receiving the correct dose. For scanning beams, generally two monitor chambers, which encompass the entire range of the scanning beam, are used to provide two independent measures of beam flux. The ionization chambers also monitor the dose delivered to a given spot. |
About OncoLink Contact OncoLink Privacy statement Disclaimer Link to OncoLink Home |